Sediplus® System Solutions
Sediplus® System Solutions
言語:
製品番号 | 90.1092 |
---|---|
Product description | Sediplus® S 2000 NX, automatic 40-channel blood sedimentation measuring device, incl. Test-Sedivettes for functional tests, suitable for S-Sedivette®, 40 measuring stations, for measuring the blood cell sedimentation rate (ESR), supply voltage: 110/230 V |
Application | for measuring the blood cell sedimentation rate (ESR) |
---|---|
Suitable for | S-Sedivette® |
Number of positions | 40 |
Interface | HL7 |
Width of product | 270 mm |
---|---|
Height of product | 305 mm |
Length of product | 355 mm |
Product category | In vitrodiagnostic device, CE |
---|---|
CE certified | CE |
Batched | no |
Minimum order qty. | 1 |
---|---|
Type of smallest subpackaging | case |
Piece(s) / inner box | 1 |
Piece (s) / outer case | 1 |
Piece(s) / pallet | 1 |
Depth of case | 510 mm |
Width of case | 430 mm |
Height of case | 350 mm |
Case volume | 0.0768 cbm |
Weight of product | 6.7 kg |
Weight of case | 9.2 kg |
EAN of inner box | 4038917927055 |
EAN case | 4038917927062 |
はい、Sediplus® S 2000 NXは、インターフェースを介した統合が可能です。
いいえ。各測定ステーションはそれぞれ個別のタイミングで充填することが出来ます。
はい、ESRはSediplus® S 2000 NX(40の測定ステーション搭載)で自動的に特定できます。
S-Sedivette®では、採血後、基本的に血沈のために十分混合する必要があります。CLSI 規格によると、測定は、採血後4時間以内に開始する必要があります。最初の血沈速度(ESR)の読み取りは、60分後に行う必要があります。
S-Sedivette®は、呼び容量の20%を構成するクエン酸三ナトリウム溶液で調製します。これにより、血沈に必要な比率(クエン酸塩1対血液4)が確保できます。
S-Sedivette®の保存期間は15カ月です。
はい、S-Sedivette®では、採血直後に完全な混合が必要で、S-Sedivette®を逆さにしてゆっくりと5回振ります。気泡とガラスビーズは、必ず反対側まで移動させるようにしてください!素早く振るのはおやめください。安全と時間の節約のために、専用の混合ユニットSarmix® M 2000をお勧めします。
S-Sedivette®は、ヴェステルグレン法に基づいてESR(赤血球沈降速度)を判定するために、開発と評価がされています。